▍ OptiTUIO Calibration Manual
Troubleshooting FAQ
Resolving runtime errors, application startup failures, network drops, and software resets.
Version: 20251014_V1-5-0_Beta_4
Last Updated: 20th June 2026
Frequently Asked Questions
OptiTUIO™ is not starting at all
- Root Cause: The required Microsoft Visual C++ redistributable package is missing.
- Resolution: Install the runtime by running
vcredist_x64.exelocated directly in thebin32orbin64directories of your extracted folder. If you still encounter failures, download the latest Visual C++ redistributable from the official Microsoft site.
OptiTUIO™ is not starting the User Interface (GUI)
- Root Cause: A corrupted license file or caching issue.
- Resolution:
- Open Windows File Explorer.
- Navigate to your app data folder by typing
%APPDATA%\OptiTUIOin the navigation path (this resolves toC:\Users\<username>\AppData\Roaming\OptiTUIO). - Delete the file named
license.otlic. - Restart the calibration tool.
- Re-import your valid license file under Help -> Install license.
No touch points are detected
- Checklist:
- Confirm your sensor status displays as active (green indicator) and scans are showing up as a blue area in the preview window.
- Verify that a screen has been created under the Screens tab.
- Ensure the screen has been assigned to the active device under the Devices tab.
- Confirm that an active output receiver (TUIO, Mouse, or Touch) is set up under the Output tab and mapped to the screen.
Technical Support Contact
If you have confirmed connection configurations and still experience anomalies, contact our technical support:
- Support Email:
optituio@interactive-asia.com